Beer-Lambert law

Also called Beer's law.

The intensity of a ray through matter falls as the exponential of minus the line integral of the attenuation, which makes CT data line integrals after a log.

The Beer-Lambert law is the law of exponential attenuation of light along a path through an absorbing medium. Each short piece of the path removes a fixed fraction of the light that reaches it, and this fraction is proportional to the length of the piece and to the local attenuation coefficient μ\mu. So the transmitted fraction is e−∫μ dse^{-\int \mu\,ds}. In chemistry, μ\mu is proportional to the concentration of a dye, and the log of the transmitted fraction measures the concentration. In computer graphics, the same factor is the transmittance of participating media such as fog or air (precomputed atmospheric scattering). In X-ray computed tomography, the log of the measured fraction is the line integral of μ\mu along the ray, and these line integrals are the Radon transform of the attenuation map. The law is a model: it holds for a monochromatic beam in which each photon is either absorbed or passes straight on. It fails when the beam has many energies or when scattered photons reach the detector.

Statement

Theorem. Let μ:[0,L]→[0,∞)\mu : [0, L] \to [0, \infty) be integrable, and let I0>0I_0 > 0. The equation

dIds(s)=−μ(s) I(s),I(0)=I0,\frac{dI}{ds}(s) = -\mu(s)\,I(s), \qquad I(0) = I_0,

has exactly one absolutely continuous solution II on [0,L][0, L] that satisfies it at almost every ss, and

I(s)=I0exp⁡(−∫0sμ(t) dt).I(s) = I_0\exp\left(-\int_0^s \mu(t)\,dt\right).

In particular the fraction that leaves the path is I(L)/I0=e−pI(L)/I_0 = e^{-p} with p=∫0Lμ(t) dtp = \int_0^L \mu(t)\,dt, and p=−ln⁡(I(L)/I0)p = -\ln(I(L)/I_0).

For a ray along a line ℓ\ell through an object with attenuation map μ(x)\mu(x), the theorem gives −ln⁡(I/I0)=∫ℓμ-\ln(I/I_0) = \int_\ell \mu, the X-ray transform of μ\mu on ℓ\ell. In the plane this is Rμ(θ,s)\mathcal{R}\mu(\theta, s).

Example. A ray passes through 2 cm of a tissue with μ=0.5 cm−1\mu = 0.5\ \mathrm{cm}^{-1} and then through 3 cm of a tissue with μ=0.2 cm−1\mu = 0.2\ \mathrm{cm}^{-1}. The line integral is p=0.5⋅2+0.2⋅3=1.6p = 0.5\cdot 2 + 0.2\cdot 3 = 1.6, and I/I0=e−1.6=0.20190I/I_0 = e^{-1.6} = 0.20190. The order of the two layers does not change the result.

In plain words

Light that goes through a tinted glass loses part of itself. A second, equal glass loses the same share of what is left, not the same amount. So the light falls by a fixed factor for each layer, and the total factor for a stack is the product of the factors of the layers. A log turns this product into a sum. The sum adds, over the path, how strongly each place absorbs. A CT scanner measures how much of the beam gets through, takes the log, and so gets this sum for each ray. The rule assumes that each photon is absorbed or goes straight through. A beam with many energies, or photons that bounce into the detector, break it.

Proof

Put M(s)=∫0sμ(t) dtM(s) = \int_0^s \mu(t)\,dt. Since μ\mu is integrable, MM is absolutely continuous on [0,L][0, L] with M′=μM' = \mu almost everywhere, and so is e±Me^{\pm M}, by the chain rule for absolutely continuous functions composed with a smooth function.

Existence: I(s)=I0e−M(s)I(s) = I_0 e^{-M(s)} is absolutely continuous, I(0)=I0I(0) = I_0, and I′(s)=−μ(s) I0e−M(s)=−μ(s) I(s)I'(s) = -\mu(s)\,I_0 e^{-M(s)} = -\mu(s)\,I(s) at almost every ss.

Uniqueness: let II be any absolutely continuous solution, and put J(s)=I(s) eM(s)J(s) = I(s)\,e^{M(s)}. A product of absolutely continuous functions on a compact interval is absolutely continuous, and at almost every ss

J′(s)=(I′(s)+μ(s) I(s)) eM(s)=0.J'(s) = \big(I'(s) + \mu(s)\,I(s)\big)\,e^{M(s)} = 0.

An absolutely continuous function with derivative 0 almost everywhere is constant, so J(s)=J(0)=I0J(s) = J(0) = I_0, and I(s)=I0e−M(s)I(s) = I_0 e^{-M(s)}. □\square

The equation itself is the physical assumption. Over a short piece dsds, a photon at ss is absorbed with probability μ(s) ds\mu(s)\,ds, independently of what happened before. The expected number of photons then satisfies I(s+ds)=I(s) (1−μ(s) ds)I(s + ds) = I(s)\,(1 - \mu(s)\,ds), which is the equation in the limit.

Consequences

  • Products of layers. For a path through layers with constant μi\mu_i and lengths ℓi\ell_i, the transmitted fraction is ∏ie−μiℓi=e−∑iμiℓi\prod_i e^{-\mu_i\ell_i} = e^{-\sum_i \mu_i\ell_i}. Ray marching in volume rendering computes transmittance by this product, one step at a time.
  • CT data are linear in the object. The log data −ln⁡(I/I0)-\ln(I/I_0) are linear in μ\mu, so the sinogram is the Radon transform of μ\mu and linear methods such as filtered back-projection apply. Without the log, the data are not linear in μ\mu.
  • Half-value layer. In a uniform medium the intensity halves over the length ln⁡2/μ\ln 2/\mu. For μ=0.5 cm−1\mu = 0.5\ \mathrm{cm}^{-1} this is 1.386 cm; for μ=0.2 cm−1\mu = 0.2\ \mathrm{cm}^{-1} it is 3.466 cm.
  • Noise. A detector counts photons. If the source sends N0N_0 photons along a ray, the count is close to a Poisson variable with mean N0e−pN_0 e^{-p}. By the delta method, the estimate −ln⁡(N/N0)-\ln(N/N_0) has a standard deviation of about 1/N0e−p1/\sqrt{N_0 e^{-p}}. For N0=104N_0 = 10^4 and p=1.6p = 1.6 the mean count is 2019 and the standard deviation of the estimate is about 0.022. Rays with large pp give few counts and noisy data, and a count of 0 has no log.
  • Emission and attenuation. In emission tomography the source is inside the body, and each emitted photon is attenuated along its path out. The data are then the attenuated Radon transform of the source.

Tightness

  • Many energies. A real X-ray tube emits a spectrum, and μ\mu depends on the energy. With spectral weights wE≥0w_E \ge 0 that sum to 1, the detector gives I/I0=∑EwEe−pEI/I_0 = \sum_E w_E e^{-p_E}, where pEp_E is the line integral at energy EE. Since exp⁡\exp is convex, Jensen's inequality gives −ln⁡(I/I0)≤∑EwE pE-\ln(I/I_0) \le \sum_E w_E\,p_E. The log data grow more slowly than the thickness. For two equal energy weights with μ=0.5\mu = 0.5 and 0.2 cm−10.2\ \mathrm{cm}^{-1}, a uniform slab of thickness LL gives −ln⁡(I/I0)/L=0.339,0.298,0.264 cm−1-\ln(I/I_0)/L = 0.339, 0.298, 0.264\ \mathrm{cm}^{-1} at L=1,5,10L = 1, 5, 10 cm, while the mean of μ\mu is 0.35. This is beam hardening: the low energies are absorbed first and the beam gets harder along the path. It causes cupping and streaks in CT images.
  • Scatter. The equation counts a scattered photon as removed. If scattered photons still reach the detector, the measured II is too large and −ln⁡(I/I0)-\ln(I/I_0) is too small. In a fog or in the air, scattering into the path adds light, and the radiative transfer equation replaces the law (participating media).
  • Large concentrations. In chemistry the law that absorbance is proportional to concentration fails at large concentrations, where the molecules interact and the absorption coefficient changes. The theorem then still holds for the actual μ\mu, but μ\mu is no longer proportional to the concentration.
  • Waves. The law treats light as rays. When the features of the medium are not large compared with the wavelength, diffraction moves energy off the ray, and the straight-line model fails.

History

Bouguer stated in 1729 that equal thicknesses of a uniform medium take away equal fractions of the light. Lambert gave the law a mathematical form in his Photometria of 1760. Beer showed in 1852 that the absorption of red light in colored solutions depends on the product of concentration and thickness, which made the law a tool of chemical analysis. In tomography, the law is the reason that the measured data, after a log, are line integrals. Cormack in 1963 and Hounsfield in 1973 used it in this way.

References

  1. P. Bouguer, Essai d'optique sur la gradation de la lumière, 1729, whole book (equal thicknesses of a uniform medium take away equal fractions of the light)
  2. J. H. Lambert, Photometria, 1760, whole book (the exponential law of absorption in mathematical form)
  3. A. Beer, Bestimmung der Absorption des rothen Lichts in farbigen Flüssigkeiten, Annalen der Physik 162(5), 1852, pp. 78 to 88 (the absorption of a solution depends on the product of concentration and thickness)
  4. A. C. Kak and M. Slaney, Principles of Computerized Tomographic Imaging, SIAM Classics in Applied Mathematics 33, 2001, Chapter 4, Section 4.1 (the attenuation of monochromatic X-rays and the line integral after a log, and the polychromatic case)

Where this leads

About this page

The build computes these levels from the evidence in the repository. Each row says what the next level needs. Read what the levels mean.

LadderLevelWhat the next level needs
Human proofH0 no proof
  • add a proof in proofs/ (06.5)
LeanL0 no Lean part
  • name a Lean module in the lean field (07)
VisualV0 none
  • add viz/scene.ts or a figure (10)
WritingW4 complete
  • a review of the text: no review record reviews/J72ZYM/text.yaml