Definitionintro
Sinogram
The Radon transform data shown as an image over angle and detector position, where each point of the object traces a sinusoid.
A sinogram is the data of parallel-beam tomography, the Radon transform , drawn as an image with the angle on one axis and the detector position on the other. Each row is one projection. A single point of the object at contributes only on the curve , which is a sinusoid in . So the image of any object is a sum of sinusoids, and this gives the name. Reconstruction methods such as filtered back-projection take a sinogram as input.
Definition
Let be an integrable function on the plane with Radon transform , where and . The sinogram of is the function on the strip , seen as an image.
A scanner measures a discrete sinogram: an array with rows and columns. Row belongs to the angle , and column to the detector bin with center , where is the detector spacing. Ideally , or an average of over the bin. Some systems store angles over ; then each line is measured twice, because .
The sinusoid of a point. Write the point in polar form, . Then
The amplitude of the sinusoid is the distance of the point from the center of rotation, and its phase is the polar angle . Points at the center give a straight line . The rows at and at are mirror images in , so the sinogram over joins its own mirror image at the edge .
Properties. Each of these holds for every integrable :
- Support. If vanishes outside the disk of radius about the origin, the sinogram vanishes for . A detector of half-width then sees the whole object.
- Constant row mass. for each , so all rows have the same sum.
- Centroid sinusoid. If and is the centroid of , then . So the centroid of each row lies on the sinusoid of the point . A detector whose center is off by adds to every row centroid, which is a simple way to find the center of rotation.
- Higher moments. The -th moment of each row is a homogeneous polynomial of degree in and (the Helgason-Ludwig conditions, Natterer, Section II.4). Measured sinograms with noise or beam hardening fail these conditions.
Examples and non-examples
One point. The point has and rad. Its sinusoid passes through these values:
| 0 | ||||
|---|---|---|---|---|
| 0.5 | 0.53033 | 0.25 | -0.17678 |
The largest value is at .
A centered disk. The disk of radius with density 1 has the projection at each angle. Its sinogram has equal rows: a band of half-width about . A disk of radius centered at gives a band of the same width that follows the sinusoid of .
The Shepp-Logan phantom. The ten ellipses of the Shepp-Logan phantom give a sinogram with a closed form, the sum of the ellipse projections. The head fills a band whose half-width goes between 0.69 at and 0.92 at , the extents of the outer ellipse along .
Sampling. A discrete sinogram with angles and bins has equal sample spacing on the outer ring of the Fourier plane when , the rule derived on the filtered back-projection page. With far fewer rows the problem becomes sparse-view CT.
Non-example: raw intensities. The detector counts are not a sinogram in this sense. The values are, by the Beer-Lambert law, and only for a monochromatic beam without scatter.
Non-example: fan-beam data. A fan-beam scanner records the data over the source angle and the fan angle. A point does not trace the sinusoid above in these coordinates, and the rows are not parallel projections. Fan-to-parallel rebinning resorts the data into a parallel sinogram.
Related definitions
- Radon transform: the function that the sinogram shows.
- Back-projection operator: the adjoint, which maps a sinogram back to an image by summing along each sinusoid.
- Hough transform: in its form, an edge point votes along the same sinusoid; the Hough accumulator of a point set is a discrete sinogram of the points.
- Metal artifact reduction: methods that find the trace of a metal object in the sinogram and replace it.
References
- A. C. Kak and M. Slaney, Principles of Computerized Tomographic Imaging, SIAM Classics in Applied Mathematics 33, 2001, Chapter 3, Section 3.1, pp. 49 to 56 (parallel projections at a set of angles)
- F. Natterer, The Mathematics of Computerized Tomography, SIAM Classics in Applied Mathematics 32, 2001, Section II.4, the ranges (the moment conditions); Chapter III, sampling and resolution
Where this leads