MRI signal equation
Also called Fourier encoding model.
The received MRI signal at time t is the integral of coil sensitivity times magnetization times exp(-i 2 pi k(t) . r): a Fourier sample at k(t), the time integral of the gradient times gammabar.
The MRI signal equation is the forward model of magnetic resonance imaging. It says that the signal a receive coil picks up at time is one sample of the Fourier transform of the image, weighted by the coil sensitivity, at the spatial frequency . The gradient coils move through k-space: is times the time integral of the gradient. The equation follows from the Bloch equations when relaxation and field errors during the readout are ignored, and it turns image reconstruction into the inversion of a Fourier transform from samples. Lauterbur (1973) encoded position with gradients, Kumar, Welti and Ernst (1975) used a Fourier transform, and Ljunggren and Twieg (1983) gave the k-space form used here.
The machine
The unknown. The transverse magnetization right after excitation, a complex function of the position ( for a slice, for a volume), zero outside the field of view.
The parameters.
- The gradient waveform , in T/m.
- The k-space trajectory , in cycles per meter, with MHz/T for the proton.
- The coil sensitivity , complex; for one uniform coil. With coils there are functions .
The forward map.
The sign is that of the forward Fourier transform and of the forward DFT. Some scanners and books use the opposite sign or demodulation; a flip of the sign gives a mirrored image. The noise is circular complex Gaussian: real and imaginary parts are independent with the same variance, white over time for one coil, and correlated between coils.
The discrete model. On a Cartesian grid with the field of view and samples per axis, the k-space spacing is and the pixel size is . With the image of pixels, the data are
where is the unitary DFT, so that , selects the acquired samples and is the diagonal matrix of the sensitivity of coil . Non-Cartesian data use the nonuniform discrete Fourier transform in place of . This page uses throughout.
Where it holds. During one readout of a few milliseconds, when is long compared with the readout, the field is uniform apart from the gradients, and nothing moves.
Where it fails, and the usual extensions.
- Relaxation during a long readout multiplies the integrand by . This blurs the image along the direction in which time grows, which matters in echo planar and spiral imaging.
- A field error multiplies it by . This shifts and blurs the image; conjugate phase reconstruction corrects it with a field map.
- Motion between or during readouts makes the image different for each sample, which PROPELLER and other methods correct.
- Gradient errors and eddy currents make the true differ from the planned one.
What it counts
- The data. One complex number per sample time and per coil.
- The resolution and field of view. With samples up to on each axis, the pixel size is . With the spacing , the image repeats with the period , which must be at least the size of the object.
- The gradient a scan needs. For mm and , cycles per meter and cycles per meter. A readout with a dwell time of 4 microseconds needs mT/m and takes 1.024 ms, a bandwidth of 977 Hz per pixel.
- The reconstruction problem. The model is the forward operator of the MRI reconstruction problem; a fit minimizes plus a penalty, which is the negative log-likelihood under Gaussian noise.
Relation to other models
- Bloch equations. In the field and without relaxation, the transverse magnetization at turns by . The coil integrates over the body, and the receiver removes the carrier . What remains is the signal equation.
- Fourier transform. For a fixed time, , the Fourier transform of the weighted image at one point. A Cartesian scan samples it on a grid, so the image is an inverse DFT away.
- Projections. A radial line through the center of k-space is the 1D Fourier transform of the projection of the image at that angle, by the Fourier slice theorem. Early MRI used projection reconstruction for this reason.
- Parallel imaging. With several coils, the sensitivities add spatial information; SENSE and GRAPPA use it to undersample.
- Dynamic models. Time-resolved MRI makes depend on time; the partial separability model restricts it to a low rank form.
Known results in the model
Conjugate symmetry (theorem). If is real, then . Half of k-space then determines the rest. In practice has a phase, and partial Fourier reconstruction uses the symmetry with a model of that phase.
Proof. when is real.
Full Cartesian sampling is invertible (theorem). If all grid samples are acquired with one coil and no noise, . With noise circular Gaussian of variance per sample, the error has the same distribution, because is unitary.
Undersampling folds the image (theorem). In 1D, let and keep every -th sample, for . Then these are the DFT of length of the folded image . So the image from the kept samples is the sum of copies of the true image, shifted by : the field of view shrinks by and the image folds times.
Proof. . Write with and . Then , and . So .
What the data cannot see. With a mask that drops samples, every image whose spectrum lives only on the dropped samples gives zero data. So the problem has a null space, and a unique answer needs prior knowledge: a smooth phase (partial Fourier), coil sensitivities (SENSE), or sparsity (compressed sensing MRI).
Least squares is the maximum likelihood estimate. With white circular Gaussian noise, the negative log-likelihood is plus a constant, where is the forward operator. For coils with a noise covariance , the weight is .
References
- Zhi-Pei Liang and Paul C. Lauterbur, Principles of Magnetic Resonance Imaging, A Signal Processing Perspective, IEEE Press, 2000, the chapter on signal localization (frequency and phase encoding and the k-space interpretation; chapter number not confirmed)
- Paul C. Lauterbur, Image formation by induced local interactions, examples employing nuclear magnetic resonance, Nature 242(5394), 1973, 190 to 191, position encoding by field gradients (general locator)
- Stig Ljunggren, A simple graphical representation of Fourier-based imaging methods, Journal of Magnetic Resonance 54(2), 1983, 338 to 343, the signal as a sample of the Fourier transform at a point that the gradients move (general locator)
- Donald B. Twieg, The k-trajectory formulation of the NMR imaging process with applications in analysis and synthesis of imaging methods, Medical Physics 10(5), 1983, 610 to 621, the k-trajectory formulation of the signal (general locator)
- Anil Kumar, Dieter Welti and Richard R. Ernst, NMR Fourier zeugmatography, Journal of Magnetic Resonance 18(1), 1975, 69 to 83, image formation by phase and frequency encoding and a Fourier transform (general locator)
Where this leads
About this page
The build computes these levels from the evidence in the repository. Each row says what the next level needs. Read what the levels mean.
| Ladder | Level | What the next level needs |
|---|---|---|
| Human proof | H0 no proof |
|
| Lean | L0 no Lean part |
|
| Visual | V0 none |
|
| Writing | W4 complete |
|
Tag T3J6N5 Edit this page History Report an error Markdown