Photoacoustic tomography

Also called optoacoustic tomography, thermoacoustic tomography.

Recover the initial pressure that a laser pulse makes in tissue from the acoustic waves recorded on a surface around it.

Photoacoustic tomography (PAT) is the problem of recovering the initial pressure p0(x)p_0(x) that a short pulse of light creates in tissue, from the sound waves that this pressure sends to detectors outside the tissue. Light absorbed in a few nanoseconds heats the tissue a little and makes it expand, so p0p_0 is proportional to the absorbed energy, which shows blood vessels and other strong absorbers with optical contrast at ultrasound resolution. When the pulse is a microwave pulse, the same problem is called thermoacoustic tomography. Mathematically, the data are the values on a surface of the solution of the wave equation with initial value p0p_0 and zero initial velocity, or equivalently the means of p0p_0 over spheres centred on that surface (Kuchment and Kunyansky 2008). With a closed surface and a constant speed of sound the problem has a unique and stable solution and exact formulas such as universal back-projection. With a partial surface, a variable speed of sound or attenuation it becomes harder.

In plain words

Flash a very short pulse of light into tissue. The parts that absorb more light, such as blood, warm up a little more and expand, and each of them becomes a small source of sound. Microphones around the tissue record the sound.

The problem is to work back from the recordings to the map of the sources. Sound from a source at distance rr reaches a microphone after the time r/cr/c. So each moment of a recording is the sum of all sources on a sphere around the microphone. With microphones all around, these sums over spheres pin down every source. This is like a CT scan, with spheres in place of straight lines.

Definition

Model. The initial pressure p0p_0 is smooth with compact support in a region Ω\Omega. The speed of sound cc is known, usually taken constant (c=1540c = 1540 m/s is the usual soft tissue convention). The pressure solves

ptt=c2Δpin Rd×(0,∞),p(x,0)=p0(x),pt(x,0)=0.p_{tt} = c^2 \Delta p \quad \text{in } \mathbb{R}^d \times (0, \infty), \qquad p(x, 0) = p_0(x), \qquad p_t(x, 0) = 0.

The zero initial velocity comes from the short pulse: the tissue is heated before it has time to move.

Data. g(z,t)=p(z,t)g(z, t) = p(z, t) for zz on a detection surface SS outside Ω\Omega and 0≤t≤T0 \le t \le T.

Task. Find p0p_0 from gg.

Spherical means. In three dimensions, p(z,t)=∂t(t Mp0(z,ct))p(z, t) = \partial_t \big(t\, M p_0(z, ct)\big), where Mp0(z,ρ)M p_0(z, \rho) is the mean of p0p_0 over the sphere of radius ρ\rho about zz. So the data determine, and are determined by, the spherical means of p0p_0 centred on SS, the spherical mean Radon transform.

Variants

  • Closed surface or partial surface. A sphere, a cylinder or a plane around the object, or only part of one (limited view).
  • Dimension. 3D with point detectors, or 2D with line detectors that integrate the pressure along a line, or a 2D slice with focused detectors. In even dimensions the wave has a tail, and the inversion formulas differ.
  • Variable speed of sound. c(x)c(x) known but not constant, as in breast imaging.
  • Attenuation and finite detector bandwidth. The data are filtered versions of gg.
  • Quantitative PAT. The next step: recover the optical absorption from p0p_0, which is the absorption times the light fluence times a conversion factor.
  • Thermoacoustic tomography. The same problem with microwave or radio frequency heating.

Examples

A uniform ball. Let p0=1p_0 = 1 in a ball of radius aa and 0 outside, in 3D. At the distance r>ar > a from its centre, the pressure is

p(r,t)=r−ct2rfor ∣r−ct∣<a,0 otherwise.p(r, t) = \frac{r - ct}{2r} \quad \text{for } |r - ct| < a, \qquad 0 \text{ otherwise}.

With a=1a = 1 mm, r=10r = 10 mm and c=1.54c = 1.54 mm/µs, the detector sees an N-shaped pulse that falls linearly from +0.05+0.05 to −0.05-0.05 between t=5.84t = 5.84 µs and t=7.14t = 7.14 µs. The amplitude falls as 1/r1/r, the duration 2a/c2a/c gives the size of the ball, and the arrival time r/cr/c gives its distance. The sharp edges of the N come from the sharp edge of the ball.

A Gaussian ball. For p0=exp⁡(−∣x∣2/(2s2))p_0 = \exp(-|x|^2/(2 s^2)), the pressure at distance rr is ((r−ct)g(r−ct)+(r+ct)g(r+ct))/(2r)\big((r - ct) g(r - ct) + (r + ct) g(r + ct)\big)/(2r) with g(u)=e−u2/(2s2)g(u) = e^{-u^2/(2s^2)}, a smooth N. The page on universal back-projection uses it to check that the reconstruction is exact.

Limited view. With detectors on only the upper half of a sphere, the edges of an object whose normals do not point toward any detector are blurred in every reconstruction. In the test of the universal back-projection page, the value at the centre of a Gaussian ball becomes 0.577 instead of 1.

Known algorithms

Method Geometry Exact Cost
Universal back-projection (Xu and Wang 2005) sphere, cylinder, plane in 3D yes in 3D O(VD)O(V D)
Inversion formulas of Finch, Patch and Rakesh (2004) sphere, odd dimensions yes O(VD)O(V D)
Formulas of Finch, Haltmeier and Rakesh (2007) sphere, even dimensions yes more than O(VD)O(V D), with an integral over time
Formulas of Kunyansky (2007) sphere yes O(VD)O(V D)
Fourier methods plane yes, for full data O(n3log⁡n)O(n^3 \log n) with the FFT and interpolation
Time reversal (Hristova, Kuchment and Nguyen 2008) any closed surface, variable cc yes in 3D with constant cc; approximate otherwise a wave solver
Delay and sum any no O(VD)O(V D)
Model based iterative methods any, with noise and filters no closed form iterative

Here VV is the number of image points and DD the number of detectors; for n3n^3 voxels and n2n^2 detectors, O(VD)=O(n5)O(V D) = O(n^5).

AlgorithmTimeSpacePropertiesCode
Universal back-projectionO(D T + V D) worstO(D T) extradeterministicexactlinear in dataiterative1 language

Bounds

  • Uniqueness (theorem). For a closed surface around the support and a constant speed, the data determine p0p_0 (Finch, Patch and Rakesh 2004; Kuchment and Kunyansky 2008 survey this and the more general cases).
  • Stability (theorem, surveyed by Kuchment and Kunyansky 2008). With a closed surface the inversion is about as stable as the inversion of the Radon transform. With limited view, the parts of the boundaries of objects that are not seen at a normal direction from some detector cannot be recovered stably.
  • Record length. In 3D with a constant speed, the pressure inside a sphere of radius RR is zero after 2R/c2R/c (Huygens' principle), so a record of that length holds all the information. In 2D the wave has a tail, and every finite record cuts it.
  • Cost. Reading the data takes Ω(DT)\Omega(D T) for TT time samples per detector.

Reductions

To the spherical mean transform. In 3D, p=∂t(t Mp0(⋅,ct))p = \partial_t (t\, M p_0(\cdot, ct)), and the map from tMp0t M p_0 to pp is invertible by integration in time, because tMp0t M p_0 is zero at t=0t = 0. So PAT with point detectors is the inversion of the spherical mean transform with centres on SS.

To the Radon transform. When the detectors are far from the object compared with its size, the spheres through the object are almost planes, and the spherical means become plane integrals: PAT tends to the inversion of the 3D Radon transform. Back-projection along spheres is then the analogue of filtered back-projection.

To a boundary value problem. Time reversal solves the wave equation backward from t=Tt = T with zero data inside and the measured gg as boundary values on SS. In 3D with a constant speed and T≥2R/cT \ge 2R/c this gives p0p_0 exactly.

Compared with diffraction tomography. In diffraction tomography a known incident wave scatters from the object. In PAT the object itself is the source. Both are inverse problems for the wave equation, but PAT is linear without any approximation, while diffraction tomography is linear only under the Born or Rytov approximation.

Applications

  • Preclinical imaging of small animals: the vascular system, tumours and the brain.
  • Breast imaging with hemispherical or ring arrays of detectors.
  • Photoacoustic microscopy of blood vessels and of the oxygen saturation of blood, which needs several wavelengths of light.
  • Thermoacoustic imaging with microwave heating, with the contrast of the electrical conductivity.
  • Hybrid ultrasound systems that use the same array for pulse echo images and for photoacoustic images.

References

  1. Peter Kuchment and Leonid Kunyansky, Mathematics of thermoacoustic tomography, European Journal of Applied Mathematics 19(2), 2008, 191, the model with zero initial velocity, the spherical mean transform, uniqueness, stability, inversion formulas and limited data (general locator)
  2. Minghua Xu and Lihong V. Wang, Photoacoustic imaging in biomedicine, Review of Scientific Instruments 77, 2006, 041101, the physics of photoacoustic generation and a review of reconstruction methods (general locator)
  3. Minghua Xu and Lihong V. Wang, Universal back-projection algorithm for photoacoustic computed tomography, Physical Review E 71, 2005, 016706, a back-projection formula exact for spherical, cylindrical and planar surfaces (general locator)
  4. David Finch, Sarah K. Patch and Rakesh, Determining a function from its mean values over a family of spheres, SIAM Journal on Mathematical Analysis 35(5), 2004, 1213, uniqueness and exact inversion formulas for spherical means centred on a sphere in odd dimensions (general locator)
  5. David Finch, Markus Haltmeier and Rakesh, Inversion of spherical means and the wave equation in even dimensions, SIAM Journal on Applied Mathematics 68(2), 2007, 392, exact inversion formulas in even dimensions (general locator)
  6. Leonid A. Kunyansky, Explicit inversion formulae for the spherical mean Radon transform, Inverse Problems 23(1), 2007, 373, explicit inversion formulas for the sphere (general locator)
  7. Yulia Hristova, Peter Kuchment and Linh Nguyen, Reconstruction and time reversal in thermoacoustic tomography in acoustically homogeneous and inhomogeneous media, Inverse Problems 24(5), 2008, 055006, time reversal reconstruction for constant and variable speed of sound (general locator)

Where this leads

About this page

The build computes these levels from the evidence in the repository. Each row says what the next level needs. Read what the levels mean.

LadderLevelWhat the next level needs
Human proofH0 no proof
  • add a proof in proofs/ (06.5)
LeanL0 no Lean part
  • name a Lean module in the lean field (07)
VisualV0 none
  • add viz/scene.ts or a figure (10)
WritingW2 basic
  • a contract.yaml