Definitionadvanced

Geometry factor

Also called g-factor.

The noise amplification of a parallel imaging method at each pixel, beyond the sqrt(R) loss from fewer samples. For SENSE it is sqrt([M^-1]_pp [M]_pp) with M = E^H Psi^-1 E, and it is at least 1.

The g-factor, or geometry factor, measures how much a parallel imaging reconstruction amplifies the noise at each pixel. Skipping samples costs a factor R\sqrt{R} in signal to noise ratio, because the scan collects RR times fewer samples. The coil geometry adds a second loss, the g-factor, because the unfolding of SENSE and the k-space fit of GRAPPA mix the noise of the coils. Pruessmann, Weiger, Scheidegger and Boesiger defined it for SENSE in 1999. It depends on the coil sensitivity maps, the noise covariance and the sampling pattern, not on the object.

Definition

Take the Cartesian SENSE model of an n×nn \times n image with cc coils, reduction factor RR, m=n/Rm = n/R, and the c×cc \times c Hermitian positive definite noise covariance Ψ\Psi between the coils. For the folded pixel (i,j)(i, j) let EE be the c×Rc \times R matrix with E[k][p]=sk(i+pm,j)E[k][p] = s_k(i + pm, j), and let

M=EHΨ−1E.M = E^H \Psi^{-1} E.

If MM is invertible, the g-factor of the image pixel (i+pm,j)(i + pm, j) is

gp=[M−1]pp [M]pp.g_p = \sqrt{[M^{-1}]_{pp}\, [M]_{pp}}.

With independent k-space noise of covariance Ψ\Psi between the coils at each sample, the SENSE output at that pixel has the variance R [M−1]ppR\,[M^{-1}]_{pp}, while the optimal combination of fully sampled data has the variance 1/[M]pp1/[M]_{pp}. So

SNRred=SNRfullgpR.\mathrm{SNR}_{\text{red}} = \frac{\mathrm{SNR}_{\text{full}}}{g_p \sqrt{R}}.

The proof on the SENSE page derives both variances and shows gp≥1g_p \ge 1 (a theorem, by the Cauchy-Schwarz inequality), with equality if and only if column pp of EE is orthogonal to the other columns in the inner product uHΨ−1wu^H \Psi^{-1} w.

For any linear reconstruction x^=Hy\hat x = H y, the same ratio defines the g-factor in general: gg at a pixel is the standard deviation of x^\hat x there, divided by R\sqrt{R} times the standard deviation of the optimal fully sampled reconstruction. It can be computed from HH (for GRAPPA, Breuer et al. 2009) or estimated by reconstructing many noise samples, the pseudo multiple replica method (Robson et al. 2008). A Monte Carlo estimate is a measurement with a sampling error.

Examples and non-examples

  • No acceleration. For R=1R = 1, MM is a 1×11 \times 1 matrix and g=1g = 1 at every pixel.
  • Orthogonal coils. If the coils of a pixel group are disjoint, for example coil 1 sees only the top half and coil 2 only the bottom half, the columns of EE are orthogonal and g=1g = 1. Then parallel imaging loses only R\sqrt{R}.
  • Two coils, R=2R = 2. With the column profiles s1=(1,0.9,0.5,0.3)s_1 = (1, 0.9, 0.5, 0.3) and s2=(0.5,0.6,1,0.9)s_2 = (0.5, 0.6, 1, 0.9), n=4n = 4 and Ψ=I\Psi = I, the pixel group {0,2}\{0, 2\} has M=(1.25111.25)M = \begin{pmatrix} 1.25 & 1 \\ 1 & 1.25 \end{pmatrix}, [M−1]00=2.2222[M^{-1}]_{00} = 2.2222 and g=1.6667g = 1.6667; the group {1,3}\{1, 3\} has g=1.6288g = 1.6288 (the example of SENSE).
  • R close to the number of coils. For four Gaussian ring coils on a 32×3232 \times 32 grid (the maps of the coil sensitivity map page), the g-factor had the mean 1.342 and the maximum 2.918 at R=2R = 2, and the mean 67.3 and the maximum 393 at R=4R = 4 (measurement with the SENSE code). With four coils, only about two of them differ along the phase encoding axis at each pixel group, so R=4R = 4 is nearly singular.
  • Non-example: the R\sqrt{R} loss. The factor R\sqrt{R} comes from the shorter scan and is outside the g-factor. A scan with g=1g = 1 still loses R\sqrt{R}.
  • Non-example: nonlinear methods. For compressed sensing MRI and other nonlinear reconstructions, the noise in the output depends on the object and on the noise level, so a g-factor map is defined only for a given object and a given noise level, and it is a measurement.
  • SENSE is the reconstruction for which the closed form holds; CG-SENSE has the same noise at convergence for the same sampling.
  • GRAPPA and slice-GRAPPA have their own g-factors, from the image domain form of their kernels. In simultaneous multi-slice imaging, the CAIPI shift lowers the g-factor by moving the aliased copies apart.
  • The root-sum-of-squares and the optimal coil combinations of fully sampled data give the reference noise level 1/[M]pp1/[M]_{pp}.
  • Noise models for MR images and the conversion to SNR units: image noise; Kellman and McVeigh (2005).

References

  1. Klaas P. Pruessmann, Markus Weiger, Markus B. Scheidegger and Peter Boesiger, SENSE, sensitivity encoding for fast MRI, Magnetic Resonance in Medicine 42(5), 1999, 952 to 962, the definition of the geometry factor and the relation SNR_red = SNR_full / (g sqrt(R)) (general locator; equation numbers not confirmed)
  2. Philip M. Robson, Aaron K. Grant, Ananth J. Madhuranthakam, Riccardo Lattanzi, Daniel K. Sodickson and Charles A. McKenzie, Comprehensive quantification of signal-to-noise ratio and g-factor for image-based and k-space-based parallel imaging reconstructions, Magnetic Resonance in Medicine 60(4), 2008, 895 to 907, the pseudo multiple replica method for g-factor maps of any linear reconstruction (general locator)
  3. Felix A. Breuer, Stephan A. R. Kannengiesser, Martin Blaimer, Nicole Seiberlich, Peter M. Jakob and Mark A. Griswold, General formulation for quantitative g-factor calculation in GRAPPA reconstructions, Magnetic Resonance in Medicine 62(3), 2009, 739 to 746, the analytic g-factor of GRAPPA from the image domain form of the kernel (general locator)
  4. Peter Kellman and Elliot R. McVeigh, Image reconstruction in SNR units, a general method for SNR measurement, Magnetic Resonance in Medicine 54(6), 2005, 1439 to 1447, SNR maps in units of the noise standard deviation, with the noise covariance and the g-factor (general locator)

About this page

The build computes these levels from the evidence in the repository. Each row says what the next level needs. Read what the levels mean.

LadderLevelWhat the next level needs
Human proofH0 no proof
  • add a proof in proofs/ (06.5)
LeanL0 no Lean part
  • name a Lean module in the lean field (07)
VisualV0 none
  • add viz/scene.ts or a figure (10)
WritingW4 complete
  • a review of the text: no review record reviews/1N3E4K/text.yaml